Health product recall

Hugo Ras Surgeon Console

Brand(s)
Last updated

Summary

Product
Hugo Ras Surgeon Console
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Hugo Ras Surgeon Console

C22AJK0267

MRASC0001

Issue

There has been 11 reported field failure incidents of power loss of main AC/DC power supply of Hugo Ras Surgeon Console. The power supply failure could lead to the permanent loss of ability to teleoperate the system from the surgeon console before or during the surgical procedure. During this failure, the tower and arm cart assembly remain operable, enabling the manual manipulation of the arms and/or the removal of the instruments and endoscope, if needed.

Recall start date: May 18, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Covidien Llc
15 Hampshire Street, Mansfield, Massachusetts, United States, 02048
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75314

Get notified

Receive notifications for new and updated recalls and alerts by category.