Hugo Ras Surgeon Console
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Hugo Ras Surgeon Console |
C22AJK0267 |
MRASC0001 |
Issue
There has been 11 reported field failure incidents of power loss of main AC/DC power supply of Hugo Ras Surgeon Console. The power supply failure could lead to the permanent loss of ability to teleoperate the system from the surgeon console before or during the surgical procedure. During this failure, the tower and arm cart assembly remain operable, enabling the manual manipulation of the arms and/or the removal of the instruments and endoscope, if needed.
Recall start date: May 18, 2024
Additional information
Details
Covidien Llc |
15 Hampshire Street, Mansfield, Massachusetts, United States, 02048 |
Get notified