This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

HP TORIC (2017-11-13)

Starting date:
November 13, 2017
Posting date:
December 8, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-65322

Affected products

HP TORIC

Reason

Improper sterilization cycle. An incorrect autoclave cycle was inadvertently selected, resulting in improper sterilization parameters.

Affected products

HP TORIC

Lot or serial number

AZ151120

Model or catalog number

AOHPT

Companies
Manufacturer

ALDEN OPTICAL

6 LANCASTER PARKWAY

NEW YORK

14086

New York