Health product recall

Horizon® Microclip™ Titanium Ligating Clips with Tape

Brand(s)
Last updated

Summary

Product
Horizon® Microclip™ Titanium Ligating Clips with Tape
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Horizon® Microclip™ Titanium Ligating Clips with Tape

More than 10 numbers, contact manufacturer.

005200

Issue

Teleflex is initiating this voluntary recall because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures. Additionally, this introduces a further risk to the patient whereby the tips of the applier may project further past the tissue to be  ligated, potentially risking injury to other distal anatomical structures.

Recall start date: May 2, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Teleflex Medical

3015 Carrington Mill Blvd., Morrisville, North Carolina, United States, 27560

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73659

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe