Health product recall

HORIZON PUMP BLOOD SET (2021-08-04)

Starting date:
August 4, 2021
Posting date:
August 24, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-76287

Last updated: 2021-08-24

Affected Products

  • A. HORIZON PUMP BLOOD SET
  • B. HORIZON PUMP IV SETS

Reason

Through receipt of customer complaints B. Braun Medical Inc. has identified there is an elevated rate of occurrence of leakage of the Outlook Pump Sets. Specifically, leakage has been reported within the Cassette portion of the Pump Sets which is used to facilitate pumping when the Cassette is loaded inside of the Outlook Infusion Pumps.

Affected products

A. HORIZON PUMP BLOOD SET

Lot or serial number
  • 61760493
  • 61763906
  • 61765128
  • 61771281
Model or catalog number

354216