This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Horizon Medical Imaging Products (2013-12-23)

Starting date:
December 23, 2013
Posting date:
March 18, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38347

Recalled Products

Horizon Medical Imaging Products

Reason

When archiving task is stopped and then started in quick succession, software defect causes incorrect study to be presented to radiologist while requested study gets erroneously deleted without being archived.

Affected products

Horizon Medical Imaging Products

Lot or serial number

All lots

Model or catalog number

HIM 11.6

HMI 11.0

HMI 11.5

HMI 11.7

HMI 11.8

HMI 11.9

HMI 4.5

HMI 4.6.1

HMI 5

Companies
Manufacturer

Mckesson Medical Imaging Company

130 - 10711 Cambie Road

Richmond

V6X 3G5

British Columbia

CANADA