Horizon Medical Imaging Management System (2014-05-21)
- Starting date:
- June 21, 2014
- Posting date:
- July 15, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-40529
Recalled products
- Horizon Medical Imaging Management System
Reason
When two or more processes attempt to modify the same image folder within fractions of a second, the write lock mechanism may fail to prevent concurrent modification of the same data. This may result in data loss within the image folder.
Affected products
A. Horizon Medical Imaging Management System
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- HMI 11.0
- HMI 11.0.3
- HMI 11.0.4
- HMI 11.0.5
- HMI 11.0.6
- HMI 11.0.6 FR EXP2
- HMI 11.0.7
- HMI 11.0.8
- HMI 11.5
- HMI 11.5.1
- HMI 11.5.2
- HMI 11.6
- HMI 11.7
- HMI 11.8
- HMI 11.9
- HMI 4.6.1
- HMI 5
- MR 12.0
Companies
- Manufacturer
-
Mckesson Medical Imaging Company
130 - 10711 Cambie Road
Richmond
V6X 3G5
British Columbia
CANADA