Health product recall

Hoffmann® II Carbon Connecting Rod

Brand(s)
Last updated

Summary

Product
Hoffmann® II Carbon Connecting Rod
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Hoffmann® II Carbon Connecting Rod

All lots.

50796060

Hoffmann® II Carbon Connecting Rod

All lots.

50796090

Hoffmann® II Carbon Connecting Rod

All lots.

50796150

Hoffmann® II Carbon Connecting Rod

All lots.

50796050

Hoffmann® II Carbon Connecting Rod

All lots.

50796040

Hoffmann® II Carbon Connecting Rod

All lots.

50796120

Hoffmann® II Carbon Connecting Rod

All lots.

50796030

Issue

Subsequent to product complaints, Stryker has confirmed that cracking of the Hoffmann® II Carbon Connecting Rods may occur following cleaning and sterilization.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies

Stryker Gmbh

Bohnackerweg 1, Selzach, Solothurn, Switzerland, 2545

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82242

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe