HistoCore PELORIS 3
Brand(s)
Last updated
Summary
Product
HistoCore PELORIS 3
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
HistoCore PELORIS 3 | 45111184 45111172 | 45.0005 |
Issue
Leica Biosystems has become aware of a leakage issue associated with the tubing in the manifold of the HistoCore PELORIS 3 instrument. Leakage may expose instrument operators and other laboratory personnel to toxic chemicals and pose a slip hazard.
Recall start date: April 2, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Pathology
Companies
Leica Biosystems Melbourne Pty Ltd. |
495 Blackburn Road, Mount Waverley, Victoria, Australia, 3149 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77259
Get notified