Hill-Rom 100 Low Bed (2015-08-27)
- Starting date:
- August 27, 2015
- Posting date:
- September 24, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-55002
Affected Products
A. Hill-Rom 100 Low Bed
Reason
The potential for a rare occurrence of a moderate injury has been identified. a laceration, cut, tear, or abrasion to the skin on the back of the patient's lower leg could result in patients with frail skin, if the leg is positioned at the opening between the thigh and foot sections or the thigh and head sections.
Affected products
A. Hill-Rom 100 Low Bed
Lot or serial number
>1000 numbers, contact mfg.
Model or catalog number
P3930A
Companies
- Manufacturer
-
Hill-Rom Inc.
1069 STATE ROUTE 46 EAST
BATESVILLE
47006-9167
Indiana
UNITED STATES