This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Hexapod Evo-RT-Couchtop (2014-04-17)

Starting date:
April 17, 2014
Posting date:
May 6, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39345

Recalled products

  1. Hexapod Evo-RT-Couchtop

Reason

The Hexapod EVO module is fixed to the base table with six screws. If the screws are not properly tightened during installation, the screws may become loose over time. If the screws become loose, there is a potential that the patient could fall off the couchtop.

Affected products

A. Hexapod Evo-RT-Couchtop

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • P10603-700
Companies
Manufacturer

Medical Intelligence Medizintechnik Gmbh

Robert-Bosch-Strasse 8

Schwabmunchen

86830

GERMANY