Hermes Posterior Stabilized Knee - Tibial Base PL (2014-05-06)
- Starting date:
- May 6, 2014
- Posting date:
- June 5, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-39801
Recalled products
Hermes Posterior Stabilized Knee - Tibial Base PL
Reason
When these devices were blistered in the clean room before going for sterilization, a mix-up occurred between parts from a batch of Hermes tibial inserts and a batch of HI-FIT tibial inserts - Ref. 15707.
Affected products
Hermes Posterior Stabilized Knee - Tibial Base PL
Lot or serial number
- Lot 148041
Model or catalog number
- 4989
Companies
- Manufacturer
-
Ceraver Les Laboratoires Osteal médical
69 Rue de La belle Etoile, CDG Cedex,
Roissy
95957
FRANCE