Health product recall

Heater-Cooler Unit HCU 40

Last updated

Summary

Product
Heater-Cooler Unit HCU 40
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Heater-Cooler Unit HCU 40 More than 10 numbers, contact manufacturer. 70105.4917 70104.4054

Issue

Maquet Cardiopulmonary GmbH determined two trends for complaints regarding the Heater Cooler Unit HCU 40: degraded internal drain hoses and leaking vacuum valves. This recall will inform users, and give instructions to replace affected parts during next annual preventive maintenance.

Recall start date: March 13, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Maquet Cardiopulmonary GmbH

Kehler Strasse 31, Rastatt, Germany, 76437

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73341