Health product recall

HeartWare Ventricular Assist System - HeartWare Battery Charger (2018-11-26)

Starting date:
November 26, 2018
Posting date:
January 8, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68704

Affected products

HeartWare Ventricular Assist System - HeartWare Battery Charger

Reason

HeartWare, now a part of Medtronic, has been notified by an outside vendor that a quantity of HVAD Battery Charger units may have been manufactured with an incorrect circuit component. This subset may exhibit one or more of the battery charging bays not fully charging the battery and/or battery charger indicators (led) not lighting.

Affected products

Produits touchés

Lot or serial number
  • BCH000550
  • BCH000552
  • BCH000555
  • BCH000557
  • BCH001053
  • BCH001071
Model or catalog number
  • 1610CA
Companies
Manufacturer
Heartware Inc.
14400 NW 60th Avenue
Miami Lakes
33014
Florida
UNITED STATES