Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42871

Recalled Products

Heartstart MRx Defibrillator/Monitor

Reason

Philips has become aware that certain Heartstart MRx Defibrillator/Monitors can separate from the Heartstart MRx housing due to breakage of mounts on the rear case. This recall affects Heartstart MRx Monitor/Defibrillators that received an M4760A handle and cap plate upgrade shipped before 2010.

Affected products

Heartstart MRx Defibrillator/Monitor

Lot or serial number

All lots.

Model or catalog number

M3535A

M3536A

Companies
Manufacturer

Philips Medical Systems

3000 Minuteman Road

Andover

01810

Massachusetts

UNITED STATES