HeartStart Intrepid Monitor/Defibrillator
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
HeartStart Intrepid Monitor/Defibrillator |
All lots. |
867172 |
Issue
Philips has identified a potential safety and compliance issue: the HeartStart Intrepid Monitor/Defibrillator's service manual does not detail electrical safety test verification methods as required by International Electrotechnical Commission(IEC) requirements for class I devices. The current HeartStart Intrepid service manual does not detail IEC class I electrical safety test verification methods. The lack of this testing does not introduce a failure, but it may prevent a failure from being detected. This could occur when an IEC class I device is tested as a class II device after a servicing event. The issue was identified as the result of a service provider's inquiry. There have been no reports of patient harm.
Recall start date: October 19, 2023
Additional information
Details
Philips Goldway (Shenzhen) Industrial Inc.