Heartstart Defibrillation Pads
Brand(s)
Last updated
Summary
Product
Heartstart Defibrillation Pads
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Heartstart Defibrillation Pads | Not applicable | M5071A M5072A |
Issue
Some electrode pads used with the hs1/onsite/home AED have been observed to experience gel separation from the foam/tin backing when peeled from the yellow plastic liner. The gel may fold onto itself, resulting in reduced surface area of gel on the pad. A pad in this condition could cause the hs1/onsite/home AED to deliver less effective or ineffective therapy to the patient due to the reduced surface contact area with the skin.
Recall start date: November 23, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Philips Medical Systems.
22100 Bothell Everett Highway, Bothell, Washington, United States, 98021
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-72333
Get notified