Heartmate Touch Communication System
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Heartmate Touch Communication System |
All Lots. |
HMT1100 |
Issue
Abbott is notifying customers of a single reported event with Heartmate Touch™ Communication System. During a clinic visit, the Heartmate Touch Communication System in one treatment room connected to the Heartmate Touch Wireless Adapter plugged into the power module in an adjacent treatment room without a user-initiated action or notification to the user. This led to the clinician adjusting settings of another patient without the clinician's awareness.
Recall Start Date: August 1, 2023
Additional information
Details
Thoratec Corporation
6035 Stoneridge Drive, Pleasanton, California, United States, 94588