Heartmate Touch Communication System
Brand(s)
Last updated
Summary
Product
Heartmate Touch Communication System
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Heartmate Touch Communication System | All Lots. | HMT1100 HMT1100-R |
Issue
Abbott is notifying customers of a single reported event with Heartmate Touch™ Communication System. During a clinic visit, the Heartmate Touch Communication System in one treatment room connected to the Heartmate Touch Wireless Adapter plugged into the power module in an adjacent treatment room without a user-initiated action or notification to the user. This led to the clinician adjusting settings of another patient without the clinician's awareness.
Recall Start Date: August 1, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Thoratec Corporation
6035 Stoneridge Drive, Pleasanton, California, United States, 94588
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74125