Health product recall

HDS (2016-02-10)

Starting date:
February 10, 2016
Posting date:
March 4, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57408

Affected Products

HDS

Reason

A combination of events and actions was identified which leads to the problem that future automated heat disinfections may be skipped without notice to the operator. If additional condition is true, heat disinfection may be processed without permission during standby.

Affected products

HDS

Lot or serial number

All serial numbers.

Model or catalog number
  • 01HDS1800C
  • 01HDS1801C
Companies
Manufacturer
DWA GMBH & Co. KG
GROSSER SAND 8
UBSTADTER-WEIHER
GERMANY