Health product recall

Hanaulux Sterilizable Handle Hub (2019-12-17)



Starting date:
December 17, 2019
Posting date:
January 20, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72117


Last updated:

Affected Products

Hanaulux Sterilizable Handle Hub

Reason

When inserting or removing the "Hanaulux" sterilizable handle to its support, the locking mechanism may break apart, if the c-clip that retains the metal axis is broken. Parts can fall if this happens: fragments of the c-clip, the locking mechanism or parts of it, e.g. spring, metal axis. During investigation, corrosion was observed on broken circlips. It was not possible to pin down the exact root cause for the corrosion, but the most probable root cause is the use of very aggressive cleaning products.

Affected products

Hanaulux Sterilizable Handle Hub

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • ARD568303905
  • ARD568330951
  • ARD568330953
  • ARD568334999
  • ARD568335999
  • ARD568350904
  • ARD568350905
  • ARD568350951
  • ARD568350953
  • ARD568350958
  • ARD568350959
  • ARD568370955
  • ARD568370958
  • ARD568370959
Companies
Manufacturer

Maquet S.A.

Avenue de la Pomme De Pin, Parc De Limere

Ardon, Orleans

45074

FRANCE