This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Hamilton - T1 Ventilator

Starting date:
December 24, 2012
Posting date:
January 21, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Product Safety
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-16881

Recalled Products

  1. Hamilton - T1 Ventilator

Reason

Unexpected high internal oxygen consumption of Hamilton-T1 Ventilators during ventilation of small pediatric patients.  If available oxygen supply depleted, life of patient may be endangered.

Affected products

A. Hamilton - T1 Ventilator

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

160005

Companies
Manufacturer

Hamilton Medical AG