Hamilton-G5 Ventilator(software 2.40 and 2.41) (2015-02-23)
- Starting date:
- February 23, 2015
- Posting date:
- March 20, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-52601
Recalled Products
A) Hamilton-G5 Ventilator(software 2.40 and 2.41)
Reason
Ventilator display can freeze. When the Dynamic Lung is displayed, an error in the internal graphics process can occur causing the display to freeze. The patient's ventilator continues but the screen is no longer updated and the user can no longer operate the device.
Affected products
A) Hamilton-G5 Ventilator(software 2.40 and 2.41)
Lot or serial number
software 2.40 / 2.41
Model or catalog number
159001