Hamilton-G5 ventilator (2015-03-14)
- Starting date:
- March 14, 2015
- Posting date:
- March 29, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Healthcare Professionals, Hospitals, General Public
- Identification number:
- RA-57690
Affected products
A. Hamilton-G5 ventilator
Reason
Investigations have shown that after performing the suctioning maneuver, including disconnecting the patient, suctioning, and reconnecting the patient, the preset pattern of ventilation may not continue as expected.
Affected products
A. Hamilton-G5 ventilator
Lot or serial number
All lots
Model or catalog number
- 159001
- 159002
Companies
- Manufacturer
-
Hamilton Medical AG
Via Crush 8
Bonaduz
7402
SWITZERLAND