This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

HAMILTON-C2 (2017-09-08)

Starting date:
September 8, 2017
Posting date:
October 6, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-64786

Affected products

HAMILTON-C2

Reason

After a usage of two and a half years, the battery fuel gauge may indicate a higher battery charge compared to what is available. Operator is alerted by high priority alarms when batteries are depleted. If not immediately connected to an AC power outlet, the ventilator will switch to ambient mode.

Affected products

HAMILTON-C2

Lot or serial number

Not indicated

Model or catalog number
  • 160001
  • 160002
Companies
Manufacturer

Hamilton Medical AG

Via Crush 8

Bonaduz

7402

SWITZERLAND