Health product recall

Haemophilus Test Medium (2017-12-01)

Starting date:
December 1, 2017
Posting date:
January 12, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65548

Affected Products

Haemophilus Test Medium

Reason

Haemophilus Test Medium (HTM) MP0806 (Lot # 2220891) may not provide acceptable microbiological performance over time. Although the product passed initial QC, repeat QC testing has revealed a decrease in recovery of Haemophilus Influenzae ATCC: 49766, 10211 and 49247

Affected products

Haemophilus Test Medium

Lot or serial number

2220891

Model or catalog number

MP0806

Companies
Manufacturer
Oxoid Company
1926 Merivale Road Suite 100
Nepean
K2G 1E8
CANADA