This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

H3+ (PLUS) HOLTER RECORDER (2015-11-03)

Starting date:
November 3, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56084

Affected Products

H3+ (PLUS) HOLTER RECORDER

Reason

An issue with the placement of a capacitor in the H3+ Digital Holter 48 hour recorders has resulted in reduced battery life, causing the recorder to record for less than the expected time. H3+ Digital Holter recorders with serial numbers between 115110248400 and 115410276012 fail to meet the claim of 24-48 hours recording time.

Affected products

H3+ (PLUS) HOLTER RECORDER

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • BURH3PR
  • H3PLUS-BAD-XXXXX
Companies
Manufacturer
Mortara Instrument Inc.,
7865 North 86th St.,
Milwaukee
53224
Wisconsin
UNITED STATES