Health product recall

H3+ (PLUS) HOLTER RECORDER (2015-11-03)

Starting date:
November 3, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56084

Affected Products

H3+ (PLUS) HOLTER RECORDER

Reason

An issue with the placement of a capacitor in the H3+ Digital Holter 48 hour recorders has resulted in reduced battery life, causing the recorder to record for less than the expected time. H3+ Digital Holter recorders with serial numbers between 115110248400 and 115410276012 fail to meet the claim of 24-48 hours recording time.

Affected products

H3+ (PLUS) HOLTER RECORDER

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • BURH3PR
  • H3PLUS-BAD-XXXXX
Companies
Manufacturer

Mortara Instrument Inc.,

7865 North 86th St.,

Milwaukee

53224

Wisconsin

UNITED STATES