Gyrus ACMI® Classic Series Continuous Flow Resectoscope
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Classic Series Continuous Flow Resectoscope |
More than 10 numbers, contact manufacturer. |
188A |
Classic Series Continuous Flow Resectoscope |
More than 10 numbers, contact manufacturer. |
188 |
Classic Series Continuous Flow Resectoscope |
More than 10 numbers, contact manufacturer. |
194 |
Issue
This recall action is being taken because it has been determined that some of the sets of sterilization parameters provided in the IFU may not consistently render the devices adequately sterile.
Additional information
Details
Gyrus Acmi, Inc
9600 Louisiana Ave. North, Brooklyn Park, Minnesota, United States, 55445