GUNTHER TULIP RETRIEVAL SET (2021-04-08)
- Starting date:
- April 8, 2021
- Posting date:
- April 22, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75439
Last updated:
Affected Products
GUNTHER TULIP RETRIEVAL SET
Reason
Cook Inc. has identified that these lots may contain an 0.038" wire guide instead of the specified 0.035" wire guide, resulting in an incompatibility with the supplied dilator (dilator cannot be advanced over the wire guide).
Affected products
GUNTHER TULIP RETRIEVAL SET
Lot or serial number
13477341
Model or catalog number
GTRS-200-RB
Companies
- Manufacturer
-
COOK, INC
750 DANIELS WAY
BLOOMINGTON
47404
Indiana
UNITED STATES