Health product recall

Grab 'N Go Opti (2019-08-15)

Starting date:
August 15, 2019
Posting date:
November 22, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71668



Last updated: 2019-11-22

Affected Products

Grab 'N Go Opti

Reason

The recall was initiated in response to a series of similar customer complaints regarding limited functionality of the control knob.  Specifically, the control knob could not be rotated to all of the available flow settings.  Through inspection of returned units, it was determined that the limited access to flow settings was a result of the control knob having been rotated beyond its functional range.   When rotated clockwise to aux, or counterclockwise to off, the control knob is designed to engage a detent plate to prevent further rotation (the stop feature).  When forcibly rotated past the stop feature, the knob can no longer access all of the designed flow settings.

Affected products

Grab 'N Go Opti

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • PRX-9615
  • PRX-9646
Companies
Manufacturer

Western/Scott Fetzer Co.

875 Bassett Rd.

Westlake

44145

Ohio

UNITED STATES