Global PET-CT Tables
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Global Pet-Ct Table |
More than 10 numbers, contact manufacturer. |
5101900-20 |
Gt1700/Gt1700v/Vt1700/Gt2000/Vt2000 |
More than 10 numbers, contact manufacturer. |
5122080-11 |
Issue
During a specific servicing activity, if the wrong screws are removed by the service personnel without the appropriate table supports in place, there is the potential for the table to drop.
Recall start date: June 6, 2022
Additional information
Details
GE Healthcare Japan Corporation
7-127, Asahigaoka, Tokyo, Japan, 191-8503
Get notified
Receive notifications for new and updated recalls and alerts by category.