Health product recall

Giraffe™ OmniBed™ Carestation™

Last updated

Summary

Product
Giraffe™ OmniBed™ Carestation™
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Giraffe™ OmniBed™ Carestation™

More than 10 numbers, contact manufacturer.

2082844-001

Giraffe™ OmniBed™

More than 10 numbers, contact manufacturer.

2074920-001-XXXXXX

Issue

GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Datex-Ohmeda, Inc.

3114 N Grandview Blvd, Waukesha, Wisconsin, United States, 53188

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82160

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe