Health product recall

Giraffe™ OmniBed Carestation CS1

Last updated

Summary

Product
Giraffe™ OmniBed Carestation CS1
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Giraffe™ OmniBed Carestation CS1

More than 10 numbers, contact manufacturer.

2082844-001

OmniBed

More than 10 numbers, contact manufacturer.

2074920-001

Issue

GE Healthcare has become aware that for certain Giraffe™ OmniBed and Giraffe™ OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification.  This may lead to the doors becoming loose.  If the doors become loose, the system activates a high priority alarm, and the canopy movement will stop.

Recall start date: September 17, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Datex-Ohmeda, Inc.

9900 Innovation Drive, Wauwatosa, Wisconsin, United States, 53226

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76201

Get notified

Receive notifications for new and updated recalls and alerts by category.