Giraffe™ OmniBed Carestation CS1
Brand(s)
Last updated
Summary
Product
Giraffe™ OmniBed Carestation CS1
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Giraffe™ OmniBed Carestation CS1 | More than 10 numbers, contact manufacturer. | 2082844-001 |
| OmniBed | More than 10 numbers, contact manufacturer. | 2074920-001 |
Issue
GE Healthcare has become aware that for certain Giraffe™ OmniBed and Giraffe™ OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose. If the doors become loose, the system activates a high priority alarm, and the canopy movement will stop.
Recall start date: September 17, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Datex-Ohmeda, Inc.
9900 Innovation Drive, Wauwatosa, Wisconsin, United States, 53226
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76201
Get notified
Receive emails about new and updated recall and safety alerts.