Giraffe™ OmniBed Carestation CS1
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Giraffe™ OmniBed Carestation CS1 |
More than 10 numbers, contact manufacturer. |
2082844-001 |
OmniBed |
More than 10 numbers, contact manufacturer. |
2074920-001 |
Issue
GE Healthcare has become aware that for certain Giraffe™ OmniBed and Giraffe™ OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose. If the doors become loose, the system activates a high priority alarm, and the canopy movement will stop.
Recall start date: September 17, 2024
Additional information
Details
Datex-Ohmeda, Inc.
9900 Innovation Drive, Wauwatosa, Wisconsin, United States, 53226
Get notified