Health product recall

Giraffe Warmer, Panda Warmer (2019-03-21)

Starting date:
March 21, 2019
Posting date:
April 5, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69498

Last updated:

Affected Products

  1. Giraffe Warmer
  2. Panda Warmer

Reason

GE Healthcare has become aware that the bedside panels, and latch areas on the warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle. If an infant impacts a bedside panel with a cracked or broken latch, the panel can disengage and fall open, no longer protecting the infant from falling from the warmer.

Affected products

A. Giraffe Warmer

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number
  • M1118179
Companies
Manufacturer
Datex-Ohmeda, Inc
9900 Innovation Drive
Wauwatosa
53226
Wisconsin
UNITED STATES

B. Panda Warmer

Lot or serial number

More than 1000 numbers, contact manufacturer.

Model or catalog number
  • M1112198
Companies
Manufacturer
Datex-Ohmeda, Inc
9900 Innovation Drive
Wauwatosa
53226
Wisconsin
UNITED STATES