Health product recall

Giraffe Omnibed Carestation Cs1

Last updated

Summary

Product
Giraffe Omnibed Carestation Cs1
Issue
Medical devices - Device compatibility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Giraffe Omnibed Carestation Cs1

More than 10 numbers, contact manufacturer.

2082844-001

Issue

There is the potential that an incorrect type of secondary latch could have been installed on the northeast (ne) and/or northwest (nw) bedside panels for certain Giraffe Omnibed Carestation devices (serial number specific). With the incorrect type of secondary latch installed, the north wall can potentially disengage when canopy is raised. However, the device is designed so the panels will stay closed and function as intended, even if only the primary latches on the south end are properly engaged.

Recall start date: June 2, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Datex-Ohmeda, Inc.
9900 Innovation Drive, Wauwatosa, Wisconsin, United States, 53226
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73766

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe