This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Getinge Washer Disinfector 86 Series (2015-09-11)

Starting date:
September 11, 2015
Posting date:
October 5, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55208

Affected Products

Getinge Washer Disinfector 86 Series

Reason

An open port on the network interface of the 86-Series Washer Disinfector can potentially lead to a product vulnerability allowing unauthenticated remote access to the product over the network. The remote access may allow control over the internal variables and program running the machine and the disinfection processes depending on the risk for intrusion at the customer network.

Affected products

Getinge Washer Disinfector 86 Series

Lot or serial number
  • WAA056360
  • WAA056417
  • WAA057002
  • WAA057551
Model or catalog number
  • 8666
  • 8668
Companies
Manufacturer

Getinge Disinfection AB

LJUNGADALSGATAN 11

VAXJO

35115

SWEDEN