Health product recall

Getinge Washer-Disinfector (2019-04-25)

Starting date:
April 25, 2019
Posting date:
May 10, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69880


Last updated:

Affected Products

Getinge Washer-Disinfector

Reason

On some units of the model Washer-Disinfector 8668 the pump hose can collapse during the washing process. This error is a result from a faulty rubber compound in the process from supplier. If the issue occurs there will be low water pressure to the spray arms which will lead to slow movement of the spray arm and less water flow. This can result in uncleaned goods after finished process.

Affected products

Getinge Washer-Disinfector

Lot or serial number
  • WAA088620
  • WAA088622
  • WAA088692
Model or catalog number
  • 8668
Companies
Manufacturer

Getinge Disinfection AB

Ljungadalsgatan 11

Växjö

35115

SWEDEN