Getinge Washer-Disinfector (2019-04-25)
- Starting date:
- April 25, 2019
- Posting date:
- May 10, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69880
Last updated: 2019-05-14
Affected Products
Getinge Washer-Disinfector
Reason
On some units of the model Washer-Disinfector 8668 the pump hose can collapse during the washing process. This error is a result from a faulty rubber compound in the process from supplier. If the issue occurs there will be low water pressure to the spray arms which will lead to slow movement of the spray arm and less water flow. This can result in uncleaned goods after finished process.
Affected products
Getinge Washer-Disinfector
Lot or serial number
- WAA088620
- WAA088622
- WAA088692
Model or catalog number
- 8668
Companies
- Manufacturer
-
Getinge Disinfection AB
Ljungadalsgatan 11
Växjö
35115
SWEDEN