Getinge 88-Series Washer/Disinfector (2020-01-16)
- Starting date:
- January 16, 2020
- Posting date:
- January 31, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72225
Last updated: 2020-01-31
Affected Products
Getinge 88-Series Washer/Disinfector
Reason
The washer disinfectors affected by this field action can develop cracks in spot welds inside the chamber. The risk for users is due to localization infection from contaminated water (from pre-rinse) ending up on the floor although of from slipping. Although this risk is already considered mitigated with provided information in the IFU on requirements.
Affected products
Getinge 88-Series Washer/Disinfector
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
88-5
Companies
- Manufacturer
-
Getinge Disinfection AB
Ljungadalsgatan 11
Växjö
35115
SWEDEN