Health product recall

Gemini System-2 Slice

Starting date:
August 24, 2012
Posting date:
September 17, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-15564

Recalled Products

A. Gemini System-2 Slice

Reason

A bolt which attaches to the Oil Accumulator within the Gantry may fail, resulting in the Oil Accumulator detaching from the Tube Housing Assembly within the Gantry.

Affected products

A. Gemini System-2 Slice

Lot or serial number

Serial Numbers: 103, 115, 144

Model or catalog number

45356 790 0891

Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.