Health product recall

Fujifilm Mobile Digital X-Ray System FDR Nano

Last updated

Summary

Product
Fujifilm Mobile Digital X-Ray System FDR Nano
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Fujifilm Mobile Digital X-Ray System FDR Nano

SN 67250049
SN 76250095
SN 96850884

DR-XD 1000

Issue

Fujifilm became aware that in rare cases, the arm base holding the tube section in DR-XD 1000 may get damaged. A protective hook will be attached to the equipment to prevent harm to patient even if the damage to the arm spreads. There have been no reported patient injuries associated with this issue.

Recall start date: February 18, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Dental
Companies
Fujifilm Corporation
26-30 Nishiazabu, Minato-Ku, Tokyo-To, Japan, 106-8620
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77038

Get notified

Receive emails about new and updated recall and safety alerts.