This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

FT4 II (FREE THYROXINE II) (2016-05-11)

Starting date:
May 11, 2016
Posting date:
May 24, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58516

Affected products

  1. FT4 II (FREE THYROXINE II)

Reason

In the software screens on cobas e 602, the expiry date of FT4 II reagent kit lot 188609 could be displayed 11/2016, instead of the correct date 10/2016 which is the expiry date of the box and the reagent kit label. When this occurs, the false expiry date is used with the reagent kits.

Affected products

A. FT4 II (FREE THYROXINE II)

Lot or serial number

188609

Model or catalog number

6437281190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY