Health product recall

FreeO2 Automated Oxygen Therapy Device Adult & Pediatric

Brand(s)
Last updated

Summary

Product
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

FreeO2 Automated Oxygen Therapy Device Adult & Pediatric

Version HW 1.2

FO2-110-00

Issue

A battery reading error may occur when certain Freeo2 HW 1.2 devices are restarted following an upgrade from software version 1.2.X to 1.4.X. The error is immediately detectable, prevents the device from being returned to service and requires installation of a resistor at position R39.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Oxynov, Inc.
725, Boulevard Lebourgneuf, Quebec, Quebec, Canada, G2J 0C4
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82735

Get notified

Receive emails about new and updated recall and safety alerts.