FRED Easyport
Brand(s)
Last updated
Summary
Product
FRED Easyport
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Industry
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
FRED Easyport |
All lots. |
0.900000 |
Issue
Schiller Ag has become aware of instances of the FRED Easyport failing the maintenance caused by defective IGBT modules. Failure of the IGBT module may result in the device only being able to deliver a limited number of shocks.
Recall Start Date: April 1, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Schiller Ag
Altgasse 68, Baar, Zug, Switzerland, 6341
Published by
Health Canada
Audience
Industry
Recall class
Type II
Identification number
RA-77313
Get notified