Health product recall

Formula Resector Cutter, Formula Aggressive Plus Cutter and Formula Barrel Bur (2018-08-08)

Starting date:
August 8, 2018
Posting date:
August 17, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67460

Affected products

  1. Formula Resector Cutter
  2. Formula Aggressive Plus Cutter
  3. Formula Barrel Bur

Reason

Stryker's sustainability solutions division has determined that affected lots of reprocessed arthroscopic shavers may display the incorrect catalog number and speed settings when connected to the generator, require manual mode for use on core generators, or display an error message on crossfire 1 and 2 generators which does not allow the device to be used.

Affected products

A. Formula Resector Cutter

Lot or serial number

8082746

Model or catalog number

375-542-000

Companies
Manufacturer

Stryker Sustainability Solutions, Inc.

1810 West Drake Dr.

Tempe

85283

Arizona

UNITED STATES



B. Formula Aggressive Plus Cutter

Lot or serial number
  • 8069955
  • 8181408
  • 8220563
Model or catalog number

375-544-000

Companies
Manufacturer

Stryker Sustainability Solutions, Inc.

1810 West Drake Dr.

Tempe

85283

Arizona

UNITED STATES



C. Formula Barrel Bur

Lot or serial number
  • 8121786
  • 8273275
Model or catalog number

375-941-012

Companies
Manufacturer

Stryker Sustainability Solutions, Inc.

1810 West Drake Dr.

Tempe

85283

Arizona

UNITED STATES