Health product recall

Flex Monoclonal Mouse Anti-Human CD20CY, Clone L26, Ready-To-Use (Dako Omnis)

Last updated

Summary

Product
Flex Monoclonal Mouse Anti-Human CD20CY, Clone L26, Ready-To-Use (Dako Omnis)
Issue
Medical devices - Performance
What to do
Customers to be notified with instructions to follow. 
Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Flex Monoclonal Mouse Anti-Human Cd20cy, Clone L26, Ready-To-Use (Dako Omnis)

41700704

GA604

Flex Monoclonal Mouse Anti-Human Cd20cy, Clone L26, Ready-To-Use (Dako Omnis)

41810164

GA604

Issue

Weak staining has been observed in a few cases where GA604 was used to stain CLL/SLL tissue samples for CD20. This weak staining has not been observed in the majority of cases where GA604 was used to test CLL/SLL tissue. As a reminder, the diagnosis of CLL/SLL is made in the context of the histologic appearance by hematoxylin and eosin (H&E) staining, with the aid of a panel of immunohistochemical markers (routinely including other B cell markers such as PAX 5 or CD79), ancillary studies (e.g. flow cytometry on peripheral blood or bone marrow), and clinical information. 

Recall start date: July 28, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Pathology
Companies
Agilent Technologies Singapore (International) Pte Ltd.

No. 1 Yishun Avenue 7, Singapore, Singapore, 768923

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-78022

Get notified