Health product recall

Filter/HME Twinstar Plus

Last updated

Summary

Product
Filter/HME Twinstar Plus
Issue
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Filter/HME Twinstar Plus

More than 10 numbers, contact manufacturer.

MP05805
MP05810
MP05815
MP05820
MP05800

Issue

Manufacturer's market surveillance activities revealed cases of increased inspiratory resistance resulting in insufficient ventilation, which may be related to formation of condensate in the filter housing. 

Device is intended exclusively for single patient use no longer than 24 hours. Excessive condensation can significantly shorten the period of use of the device and require the product to be replaced latest after 24 hours. 

Increased resistance carries the risk of desaturation for the patient. Should this hazardous situation encounter, consider the possibility that it may be related to the occlusion of the filter due to condensate. 

The current instructions for use contain warnings related to condensate. However, manufacturer has determined that the instructions can be further improved. Until the revised instructions for use are available customers may continue using the device as long as both airway pressure and volume are permanently monitored and alarm limits for each patient are suitably selected. Change of filter is required in the case of resistance increase.

Recall start date: July 24, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Draegerwerk Ag & Co. Kgaa

Moislinger Allee 53-55, Lubeck, Germany, 23542

Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-74068