Fetal Spiral Electrode
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Fetal Spiral Electrode |
Not applicable. |
9898 031 37631 |
Issue
Philips has found that the spiral tip of the Fetal Sprial Electrode (FSE) may break off during use, potentially requiring surgical intervention to remove the broken tip from the patient. Based on Philips investigation, this can occur due to over rotation during attachment or pulling the tip from the fetal scalp. The FSE may also break off if the user pulls the spiral tip from the fetal skin, increasing the risk of the spiral tip detaching from the FSE during removal.
Recall start date: November 24, 2022
Additional information
Details
Philips Medical Systems
3000 Minuteman Road, Andover, Massachusetts, United States, 01810