Fep Introcan Safety IV Catheter - Straight (No Wings)
Brand(s)
Last updated
Summary
Product
Fep Introcan Safety IV Catheter - Straight (No Wings)
Issue
Medical devices - Sterility
What to do
Contact the manufacturer if you require additional information.
Audience
Industry
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Fep Introcan Safety IV Catheter - Straight (No Wings) | 22D07G8242 | 4251890-02 |
Issue
Investigation conducted at the manufacturing site identified deviations at the catheter hub that could lead to leakage of the product.
Recall start date: August 25, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
B. Braun Melsungen AG
Carl-Braun-Str. 1, Melsungen, Germany, 34212
Published by
Health Canada
Audience
Industry
Recall class
Type II
Identification number
RA-64535
Get notified