This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Femur Hook with Support Arm

Starting date:
February 19, 2013
Posting date:
May 15, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-29255

Recalled products

A. Femur Hook with Support Arm

Reason

Within the framework of our market surveillance, we have been made aware of one complaint in which a break of a weld of the support arm occurred. Due to the break the complete support arm could fall down. The problem was caused by inaccurate execution of the weld at a supplier. It cannot be excluded that further support arms of the femur hook devices are equipped with a faulty weld which finally could lead to the failure described above.

Affected products

A. Femur Hook with Support Arm

Lot or serial number
  • Not Applicable
Model or catalog number
  • 1433.42A0
Companies
Manufacturer

Maquet GmbH

Kehler Strasse 31

Rastatt

76437

GERMANY