Health product recall

FACSLyric Flow Cytometer (2019-08-21)

Starting date:
August 21, 2019
Posting date:
October 3, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71115


Last updated:

Affected Products

FACSLyric Flow Cytometer

Reason

BD has confirmed that someBD FACSLyric flow cytometers experience higher than expected absolute counts when using BD Trucount tubes. Excessive electronic abort counts may affect the ratio of cell population events to bead events, potentially resulting in a falsely high absolute count.

Affected products

FACSLyric Flow Cytometer

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 651164
  • 651165
  • 654587
  • 659180
  • 663029
Companies
Manufacturer

Becton, Dickinson and Company BD Biosciences

2350 Qume Drive

San José

95131

California

UNITED STATES