This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

FABIUS MRI (2015-06-15)

Starting date:
June 15, 2015
Posting date:
July 9, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54162

Recalled Product

A. FABIUS MRI

Reason

Draeger has become aware of cases in which the Fabius MRI or parts of the system were attracted by the magnetic field of the MRI. If the Fabius MRI is positioned too close to the MRI, the magnetic force of the MRI can loosen parts of the Fabius MRI and cause personal injury.

Affected products

A. FABIUS MRI

Lot or serial number
  • ARZB-0004
  • ARZD-0007
  • ARZD-0008
  • ASAD-0005
  • ASAF-0001
  • ASAF-0002
  • ASCC-0112
  • ASCC-0113
  • ASCJ-0260
  • ASCK-0034
  • ASCM-0295
  • ASDD-0060
  • ASDD-0111
  • ASEB-0154
  • ASFC-0050
  • ASFD-0196
  • ASFK-0134
  • ASFM-0169
  • ASHA-0001
Model or catalog number

8607300

Companies
Manufacturer

Draeger Medical Gmbh

Moislinger Allee 53-55

Lubeck

23542

GERMANY